750-1800 nm+1600 – 3200 nm高重频调Q激光器——Q-TUNE-HR

680-2300nm波长可调谐调Q脉冲激光器
波长:750-1800 nm+1600 – 3200 nm
能量:至1mJ,重频至100kHz,20-100 cm-1线宽
应用: 微加工。 光声microspopy。 拉曼光谱与显微镜。 红外光谱。 遥感。

基本信息

750-1800 nm+1600 – 3200 nm高重频调Q激光器——Q-TUNE-HR插图

750-1800 nm+1600 – 3200 nm高重频调Q激光器——Q-TUNE-HR

  • Up to 100 kHz pulse repetition rate.
  • Hands-free, automated wavelength tuning  from 750 to 1800 nm.
  • Up to 5 W output power in IR range.
  • Can be adapted for user-supplied laser with 532 nm or 1064 nm wavelength and up to 30 W @ 532 nm input power.
  • Microprocessor controlled operation with self-optimization, self-calibration capability.
  • Optional air purging for long lifetime of OPO optics.
  • Powered from +12 VDC source or AC/DC adapter.

Q-TUNE-HR——高重频波长可调(750-1800 nm+1600 – 3200 nm

简介

Q-TUNE-HR is an Optical Parametric Oscillator (OPO) producing tunable wavelength in 750-1800 nm range with up to 100 kHz pulse repetition rate.
Up to 100 kHz pulse repetition rate make Q-TUNE-HR perfect coherent light source for micromaching, Raman spectroscopy, microscopy, remote sensing applications.
Q-TUNE-HR needs external pump source with > 350 µJ pulse energy, 5-7 ns FWHM pulse duration @ 532 nm. Pulse repetition rate can be in 10-100 kHz range, while pump power – in 3.5-30 W range. Contact us for more details about recommended pump sources.
By default OPO is optimized to produce maximum output in 1200-1800 nm range. As option, unit can be optimized for 750-950 nm range.
All laser electronics are integrated into housing of the Q-TUNE and the only external module is mains adapter that provides 12 VDC, 20-50 W power (depending on model).
OPO is controlled trough single Ethernet port via build-in web-server. There is no need to install control software – any computer or even cell phone with modern web-browser will be able to control Q-TUNE. API is also provided for integration with user devices.
In addition to tunable wavelength output, the Q-TUNE provides bypass port for access to pump laser beam.

可选拓展

  • Air purging for long lifetime of OPO optics.
  • Compact spectrometer for monitoring of OPO wavelength and linewidth.
  • Fiber coupled OPO output.

应用

  •     Micromachining.
  •     Photo-acoustic microspopy.
  •     Raman spectroscopy & microscopy.
  •     Infrared spectroscopy.
  •     Remote sensing.

参数 1)

Model

Q-TUNE-HR

-G10K

-G100K

-I10K

Wavelength range, nm

750 – 1800 nm

1600 – 3200 nm

Pulse repetition

rate2)

1 – 10 kHz

10 – 100 kHz

1 – 10 kHz

Typical conversion efficiency

>20 % (signal + idler)

Output pulse energy

>50 µJ @ 1200 nm 3)

>100 µJ @1650 nm 4)

Linewidth

20-100 cm-1

Pulse duration 5)

5-7 ns

Typical pulse-to-pulse stability 6)

2 × pump laser

Typical power drift 7)

2 × pump laser

Polarization

horizontal

Typical beam diameter 8)

3 mm

3 mm

Typical beam divergence 9)

< 3 mrad

< 5 mrad

Pump laser requirements
Pump wavelength

532 nm

1064 nm

Pulse duration

5-8 ns FWHM

Pulse energy

> 350 µJ

>1 mJ

Power

< 40 W

M2 factor

<1.5

Dimensions
Laser head ( W×L×H)

390 × 620 × 135 mm3

Power adapter (W×L×H)

50 × 125 × 33 mm3

Operating requirements

Cooling requirements

Air cooled (water-free)

Ambient temperature

15 – 30 ºC

Relative humidity (non-condensing)

10 – 80 %

Mains voltage

90-230 VAC, single phase, 47-63 Hz 13)

Average power consumption

<50 W

1. Due to continuous improvements all specifications are subject to change. Unless stated otherwise all specifications are measured at 1200 nm and max pulse repetition rate. The parameters marked typical are not specifications. They are indications of typical performance and will vary with each unit we manufacture
2. Determined by pulse repetition rate of pump source.
3. When pumped by 4 W @532 nm @ 10 kHz pulse repetition rate source. See tuning curves for pulse energies at other wavelengths.
4. When pumped by 10 W @1064 nm @10 kHz pulse repetition rate source.
5. FWHM level at 1200 nm, measured with 350 ps rise time photodiode.
6. Measured during 30 seconds of operation after warm-up.
7. Over 8 hour period after 20 minutes of warm-up, when ambient temperature variation is less than ±2 °C. Power value is calculated every 1 second.
8. Beam diameter is measured 20 cm from laser output at the 4σ level.
9. Full angle measured at the 4σ level.
10. Laser can be powered from appropriate 12 or 28 VDC power source, depending on model. Please inquire for details.

文档